Publication | Closed Access
The Comparison of Pullout Strengths of Various Pedicle Screw Designs on Synthetic Foams and Ovine Vertebrae.
33
Citations
0
References
2015
Year
Transpedicular screws redesigned with modified helical angles exhibit higher pullout strength compared to the classical transpedicular screws and can be inserted more rapidly with the same number of screwing rounds result with doubled insertion depth.