Nuklearmedizin - NuclearMedicine · 2006 · 25 citations · 0 references
MedicineNuclear MedicineRadiologyDiagnostic ImagingDiagnostic RadiologyHybrid ImagingAttenuation CorrectionComputed TomographyDental Implant ImagingNeck Cancer ImagingNeck ImagingPositron Emission Tomography18F-fluorodeoxyglucose ImagingComputed Tomography-based Attenuation CorrectionDental ImplantsMetal Artifact CorrectionMetal Artifact ReductionPet Attenuation CorrectionPet-computed TomographyPositron Emission Tomography Attenuation Correction
UNLABELLED: PET/CT imaging is particularly promising for head/neck malignancies, but dental implants lead to biased CT attenuation and PET activity values following CT-based attenuation correction (CT-AC). OBJECTIVE: Here, we implement a metal artifact correction procedure (MAR) as part of the CT-AC for PET/CT imaging. RESULTS: Phantom studies indicate a maximum quantitative bias in CT and PET of 1000 HU and 30 %, which is reduced to 230 HU and 6 %, respectively following MAR. These results were verified in selected patients. CONCLUSION: Artifacts and biases from dental implants can be reduced in PET/CT imaging by applying a simple MAR as part of the CT-AC.