The Influence of Torque Tightening on the Position Stability of the Abutment in Conical Implant-Abutment Connections

W. D. Hogg, Kris Zulauf, Jürgen Mehrhof, Katja Nelson

The International Journal of Prosthodontics · 2015 · 14 citations · 3 references

Concepts

Abstract

The influence of repeated system-specific torque tightening on the position stability of the abutment after de- and reassembly of the implant components was evaluated in six dental implant systems with a conical implant-abutment connection. An established experimental setup was used in this study. Rotation, vertical displacement, and canting moments of the abutment were observed; they depended on the implant system (P = .001, P < .001, P = .006, respectively). Repeated torque tightening of the abutment screw does not eliminate changes in position of the abutment.

References

3