Failure to detect extremely high levels of serum IgE with an immunoradiometric assay.

J A Saryan, Patricia E. Garrett, S. R. Kurtz

PubMed · 1989 · 17 citations · 0 references

Concepts

Abstract

Two patients with extreme elevations in serum IgE were found to have falsely low levels of IgE when measured by a commercial immunoradiometric assay. Dilution studies performed on serum from both patients revealed the rise and subsequent fall of the dose-response curve characteristic of the high-dose "hook" or prozone effect. Specimens with unexpectedly low levels of serum IgE obtained by either commercial immunoradiometric assay or enzyme immunoassay should be reanalyzed after dilution.