Tc-99m human serum albumin: a suitable agent for plasma volume measurements in man.

Shirley S. L. Yang, Eileen Nickoloff, Patricia A. McIntyre, Willis C. Maddrey, Helen H. Mikesell, Ursula Scheffel, Wayne Kasecamp, Niall P. MacAllister

PubMed · 1978 · 23 citations · 3 references

Concepts

Abstract

In 16 patients we have carried out simultaneous plasma-volume measurements with human serum albumins tagged with Tc-99m (Tc-99m HSA) and I-131 (I-131 HSA). The correlation coefficient was 0.987. Tc-99m HSA, prepared from kits that predictably yield high labeling efficiency (and thereby negligible amounts of TCO4-), is clearly a superior agent for repeat plasma volume determinations, because of its shorter half-life and the reduced radiation dose to the subject.