The International Journal of Artificial Organs · 1994 · 10 citations · 9 references
MedicineCardiologySurgeryInterventional CardiologySurgical InstrumentationVascular SurgeryCardiac CatheterizationCatheter ManagementCatheter FractureEmbolized Device RetrievalForeign Body Retrieval From The Right Ventricle
A case of spontaneous fracture of the outlet catheter of a totally implanted catheter system (Port-A-Cath) is presented. The outlet catheter was fractured at the entrance into the left subclavian vein twenty-one weeks after insertion and the distal part was embolized in the right ventricle. The embolized catheter fragment was retrieved by a 'goose-neck' snare via the right femoral vein. The awareness of a possible spontaneous fracture of the outlet catheter of a totally implanted catheter system (Port-A-Cath) is important to prevent accidental spillage of potent cytotoxic substances.
9
J. David Richardson, Frederick L. Grover, J. Kent Trinkle · The American Journal of Surgery · 1974 · 149 citations