PubMed · 1994 · 13 citations · 0 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryOrthopedic BiomechanicsOrthopedicsPediatric OrthopaedicsBone BiomechanicsBone MechanicsMetatarsal BonesMetatarsal BiomechanicsMetatarsal DiaphysisMetatarsal MetaphysesScrew Fixation MechanicsTensile LoadsTensile StrengthTensile Testing
Holding power was determined for various orthopedic screws in bones of calves. Holding power was defined as maximal tensile force required to remove a screw divided by thickness of bone engaged by the screw (kN/mm). Comparative pull-out tests were performed, using pairs of large metacarpal or metatarsal bones from calves aged 3 to 14 days. Comparisons were made of the holding power of 6.5-mm fully threaded cancellous screws and 5.5-mm cortical screws in the proximal and distal metaphyses, and of 4.5-mm and 5.5-mm cortical screws in the diaphysis. Sixteen repetitions of each comparative trial were performed. There was no statistically significant difference in the holding power of 4.5- and 5.5-mm cortical screws in the diaphysis. There was no significant difference in the holding power of 5.5-mm cortical and 6.5-mm fully threaded cancellous screws in the proximal metaphysis. In the distal metaphysis, 6.5-mm fully threaded cancellous screws had significantly (P < 0.001) greater holding power than did 5.5-mm cortical screws. There was no significant difference between the mean holding power of 5.5-mm cortical screws in the proximal metaphysis and 5.5-mm cortical screws in the distal metaphysis. There was significantly (P < 0.01) greater mean holding power of 6.5-mm cortical, fully threaded cancellous screws in the distal metaphysis, compared with the proximal metaphysis.