DentistryImplant DentistryOrthopaedic SurgeryBone AugmentationImplant SurgeryOrthopedicsImplant DesignImplant Site AssessmentOsseointegration MonitoringBiomechanical Testing Of ImplantsImplant Failure MechanismsImplant Safety EvaluationOsseointegration Failure Detection
Reverse torque to failure on implants has been studied in animals. Three implants were reverse torqued to failure in a human volunteer, with failure rates between 45 and 58 Ncm. Clinical data are presented on 404 implants examined after reverse-torque testing and loading, with no increase in failure rates. Reverse-torque testing at 20 Ncm appears to be a safe, reliable method for verifying osseointegration with pure titanium screw-shaped implants.