Single-sweep-based methods to improve the quality of auditory brain stem responses Part I: Optimized linear filtering

Michael Granzow, Helmut Riedel, Birger Kollmeier

2001 · 24 citations · 17 references

Concepts

Abstract

This study is a systematic evaluation of the influence of different filter settings on the quality of auditory brain stem responses performed on a single sweep basis, i.e., on a post-hoc analysis of all unaveraged single epochs. For purposes of comparison, a quality analysis was also carried out when only two sub-averages were known. The question of optimal lower and upper filter edge frequencies is addressed for monaural and binaural stimuli at 20, 40 and 60 dB (nHL) for different recording sites. Only linear (finite impulse response) filters were used. Filter edge frequencies of 30-50 Hz and 1300-1700 Hz are to be recommended for practical purposes. These values are independent of the stimulation level and recording site. In comparison with average-based quality estimates, the single-sweep approach was found to be superior, as is confirmed both by the practical examples and by theoretical observations.

References

17