Case Studies in Nondestructive Testing and Evaluation · 2016 · 10 citations · 10 references
Veterinary MedicineRadiologyDiagnostic ImagingDiagnostic RadiologyDigital ImagingComputed TomographyComputed Tomography PhysicsImage QualityImage Quality AssessmentMedical ImagingRadiographic ImagingComputed Tomography AcquisitionComputed Tomography ArtefactsImage Performance MetricsImage Quality MetricsMedical Image Quality AssessmentObjective Image Quality MetricsRadiographic Image QualityTomographic Image Quality
This paper presents a comparison of surface-based and image-based quality metrics for dimensional X-ray computed tomography (CT) data. The chosen metrics are used to characterize two key aspects in acquiring signals with CT systems: the loss of information (blurring) and the adding of unwanted information (noise). A set of structured experiments was designed to test the response of the metrics to different influencing factors. It is demonstrated that, under certain circumstances, the results of both types of metrics become conflicting, emphasizing the importance of using surface information for evaluating the quality dimensional CT data. Specific findings using both types of metrics are also discussed.
10
Computed tomography for dimensional metrology
Jean‐Pierre Kruth, Markus Bartscher, Simone Carmignato et al. · CIRP Annals · 2011 · 672 citations
Engineering, Manufacturing Engineering, Biomedical Imaging +11
Saul N. Friedman, George S. K. Fung, Jeffrey H. Siewerdsen et al. · Medical Physics · 2013 · 199 citations