Optics Express · 2016 · 33 citations · 10 references
Natural SciencesPhysicsDiagnostic ImagingOpticsOptoelectronicsImagingInterferometryLens ImagingLens SystemsPhysical OpticsBeam TiltInterferometer DesignInterferometric ImagingInterferometric MeasurementMach-zehnder InterferometersOptical Interferometers
The coupling between beam tilt and longitudinal path length readout in a setup representing a LISA test mass interferometer was reduced to below 2 µm/rad using a two lens imaging system. This was achieved by the use of a homodyne equal arm-length Mach-Zehnder interferometer and suppression of the dominating effects of higher order Gaussian modes and longitudinal actuator movement. The latter was subtracted using the phase signal of a large single element photo diode.
10
Automatic alignment of optical interferometers
E. Morrison, B. J. Meers, D. I. Robertson et al. · Applied Optics · 1994 · 271 citations
<i>LISA</i> technology and instrumentation
O. Jennrich · Classical and Quantum Gravity · 2009 · 180 citations · Full text
The LTP interferometer and phasemeter
Gerhard Heinzel, Vinzenz Wand, O. Jennrich et al. · Classical and Quantum Gravity · 2004 · 151 citations · Full text