Concepedia

Publication | Closed Access

Bacterial endotoxin detection in hyaluronic acid-based medical devices

16

Citations

8

References

2016

Year

Abstract

A simple and rapid method has been developed for testing bacterial endotoxin in hyaluronic acid (HA)-based medical devices. High-molecular-weight HA (HMW HA) in solution or HA-based medical devices was digested by the enzyme hyaluronidase to reduce solution viscosity by truncating the long chains of HA and to test for bacterial endotoxin. The bacterial endotoxin level was detected and measured by kinetic chromogenic Limulus Amebocyte Lysate (LAL) assay. The method was applied to two different ophthalmic viscosurgical devices (OVDs) and one dermal filler, and may easily be adapted to use with up to 3% HA solutions and other HA-based medical devices. Published 2016. This article is a U.S. Government work and is in the public domain in the USA. J Biomed Mater Res Part B: Appl Biomater, 2016. © 2016 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 105B: 1210-1215, 2017.