Journal of Bone and Joint Surgery - British Volume · 1992 · 150 citations · 0 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryElbow SurgeryOrthopedic BiomechanicsOrthopedicsArthroplasty BiomechanicsElbow AnatomyElbow ArthroplastyLower-limb ProstheticsUpper-limb ProstheticsBiomechanical Load TransferBiomechanics Of Elbow ProsthesesBone-implant InterfaceConstrained Elbow ArthroplastyConstrained ProsthesesElbow BiomechanicsElbow Joint MechanicsElbow Kinematics
We used 11 cadaver elbows and a three-dimensional electromagnetic tracking device to record elbow movements before and after implantation of a 'loose-hinged' elbow prosthesis (modified Coonrad). During simulated active motion there was a maximum of 2.7 degrees (+/- 1.5 degrees) varus/valgus laxity in the cadaver joints. This increased slightly after total elbow arthroplasty to 3.8 degrees (+/- 1.4 degrees). These values are lower than those recorded for the cadaver joints and for the prostheses at the limits of their varus/valgus displacements, indicating that both behave as 'semi-constrained' joints under physiological conditions. They suggest that the muscles absorb some of the forces and moments that in a constrained prosthesis would be transferred to the prosthesis-bone interface.