Computer Methods in Biomechanics & Biomedical Engineering · 2016 · 33 citations · 24 references
DentistryImplant DentistryMechanical Behavior Of MaterialsMechanical Properties Of Dental MaterialsSurface Properties Of Dental MaterialsBone BiomechanicsBone MechanicsComputer-aided Implant DesignImplant DesignImplant MaterialsBone Strain Around ImplantsCortical Bone StructureDental Implant BiomechanicsFinite Element Analysis Of BoneFinite Element Modeling Of BoneFinite Element Strain AnalysisImplant Stress DistributionStress Transmission
The aim of this study was to assess stress/strain of different implant modeling simplifications by 3D-FEA. Three variation of external hexagon implant (Ø3.75 × 10 mm) supporting one molar crown were simulated: A (no threads); B (slightly threads simplification); C (original design). 200 N (axial) and 100 N (oblique) were applied. Cortical bone was evaluated by maximum principal stress and microstrain qualitatively and quantitatively (ANOVA and Tukey post hoc (p < 0.05)). Higher stress levels (p < 0.05) were observed in model A. Models B and C presented similar stress transmission. It was possible to conclude that slightly simplification should be used for studies evaluating stress transferring for bone tissue.
24
Eriko Kitamura, Roxana Stegaroiu, Shuichi Nomura et al. · Clinical Oral Implants Research · 2004 · 254 citations
Valentim Adelino Ricardo Barão, Juliana Aparecida Délben, J. Lima et al. · Journal of Biomechanics · 2013 · 142 citations · Full text