HemodialysisEquilibrium DialysisRia-ii ResultsRenal FunctionConcise CommunicationMedicineDisparate EffectsChronic Kidney DiseaseImmunologyThyroid DiseaseLaboratory MedicineAutoimmunityDialysis TherapyClinical ChemistryThyroid HormonePharmacologyNephrologyFree Thyroxine
Free thyroxine (FT4) has been measured in two groups of patients, before and after heparin, using equilibrium dialysis (ED) and two radioimmunoassays, RIA-I and RIA-II. In Group A, nine patients were tested before and after 12-24 hr of intravenous heparin, FT4 (ED) rose from 1.9 +/- 0.5 to 2.8 +/- 3.0 ng/dl, and RIA-I from 1.37 +/- 0.37 to 1.89 +/- 1.21 ng/dl, whereas RIA-II results fell from 0.97 +/- 0.38 to 0.66 +/- 0.32 ng/dl. In Group B, ten patients were tested before and after 15 min of intravenous heparin. FT4 (ED) rose from 1.7 +/- 0.7 to 3.2 +/- 1.6 ng/dl (p less than 0.02), and RIA-I rose from 1.3 +/- 0.46 to 2.02 +/- 0.01 ng/dl (P less than 0.05), whereas RIA-II results fell from 1.07 +/- 0.38 to 0.63 +/- 0.31 ng/dl (P less than 0.02). The correlation coefficient between FT4 (ED) and RIA-I in 38 paired results was 0.96, but there was no correlation between FT4 (ED) and RIA-II results.