Expert Review of Medical Devices · 2016 · 18 citations · 64 references
EngineeringMechanical EngineeringHeat TransferThermal AnalysisThermal ComfortCooling Rate ControlFever ManagementTemperature ManagementTemperature RegulationBurn Critical CareCoolingCooling EfficiencyCooling System EfficiencyHeat Transfer MechanismsHeat Transfer MitigationPatient Temperature Control DevicesTemperature Stability
Managing core temperature is critical to patient outcomes in a wide range of clinical scenarios. Previous devices designed to perform temperature management required a trade-off between invasiveness and temperature modulation efficiency. The Esophageal Cooling Device, made by Advanced Cooling Therapy (Chicago, IL), was developed to optimize warming and cooling efficiency through an easy and low risk procedure that leverages heat transfer through convection and conduction. Clinical data from cardiac arrest, fever, and critical burn patients indicate that the Esophageal Cooling Device performs very well both in terms of temperature modulation (cooling rates of approximately 1.3°C/hour, warming of up to 0.5°C/hour) and maintaining temperature stability (variation around goal temperature ± 0.3°C). Physicians have reported that device performance is comparable to the performance of intravascular temperature management techniques and superior to the performance of surface devices, while avoiding the downsides associated with both.
64
Therapeutic Hypothermia After Cardiac Arrest
Peter T. Morley, Terry L. Vanden Hoek, Robert W. Hickey et al. · Circulation · 2003 · 1.1K citations