Magnetic Resonance in Medicine · 2016 · 135 citations · 29 references
MedicineBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyImage Quality AssessmentImage Quality AssuranceMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMedical ImagingNuclear Medicine ImagingBland-altman PlotFat Quantification ImagingIntraclass CorrelationLinear RegressionMagnetic ResonanceMagnetic Resonance PhysicsStatistical ValidationWater Phantom Measurements
Purpose To evaluate the accuracy and reproducibility of quantitative chemical shift-encoded (CSE) MRI to quantify proton-density fat-fraction (PDFF) in a fat–water phantom across sites, vendors, field strengths, and protocols. Methods Six sites (Philips, Siemens, and GE Healthcare) participated in this study. A phantom containing multiple vials with various oil/water suspensions (PDFF:0%–100%) was built, shipped to each site, and scanned at 1.5T and 3T using two CSE protocols per field strength. Confounder-corrected PDFF maps were reconstructed using a common algorithm. To assess accuracy, PDFF bias and linear regression with the known PDFF were calculated. To assess reproducibility, measurements were compared across sites, vendors, field strengths, and protocols using analysis of covariance (ANCOVA), Bland–Altman analysis, and the intraclass correlation coefficient (ICC). Results PDFF measurements revealed an overall absolute bias (across sites, field strengths, and protocols) of 0.22% (95% confidence interval, 0.07%–0.38%) and R2 > 0.995 relative to the known PDFF at each site, field strength, and protocol, with a slope between 0.96 and 1.02 and an intercept between −0.56% and 1.13%. ANCOVA did not reveal effects of field strength (P = 0.36) or protocol (P = 0.19). There was a significant effect of vendor (F = 25.13, P = 1.07 × 10−10) with a bias of −0.37% (Philips) and −1.22% (Siemens) relative to GE Healthcare. The overall ICC was 0.999. Conclusion CSE-based fat quantification is accurate and reproducible across sites, vendors, field strengths, and protocols. Magn Reson Med 77:1516–1524, 2017. © 2016 International Society for Magnetic Resonance in Medicine
29
<i>In vivo</i> characterization of the liver fat <sup>1</sup>H MR spectrum
Gavin Hamilton, Takeshi Yokoo, Mark Bydder et al. · NMR in Biomedicine · 2010 · 511 citations
Medicine, Biochemistry, Radiology +16
Relaxation effects in the quantification of fat using gradient echo imaging
Mark Bydder, Takeshi Yokoo, Gavin Hamilton et al. · Magnetic Resonance Imaging · 2008 · 389 citations · Full text
Medicine, Biophysics, Radiology +12