Publication | Open Access
Technical Report: Reference photon dosimetry data for Varian accelerators based on IROC-Houston site visit data
39
Citations
15
References
2016
Year
IROC-H has collected numerous data on Varian Linacs and the results of photon measurements from the past 15 years are presented. The data can be used as a comparison check of a physicist's acquired values. Acquired values that are well outside the expected distribution should be verified by the physicist to identify whether the measurements are valid. Comparison of values to this reference data provides a redundant check to help prevent gross dosimetric treatment errors.