QRS Detection By Template Matching Using Real-Time Correlation On A Microcomputer

STEVEN E. DOBBS, Neil M. Schmitt, HALUK S. OZEMEK

Journal of Clinical Engineering · 1984 · 60 citations · 0 references

Concepts

Abstract

A microcomputer system which detects QRS complexes in real-time is presented. The system is unique in that it implements a template matching algorithm using moving-window correlation on a microcomputer. Incoming ECG data are continuously correlated with multiple previously stored templates. Normal or arrhythmic QRS complexes are detected whenever correlation with one of the templates exceeds a preset threshold. Use of moving-window correlation eliminates the problems often encountered when performing correlation based upon fiducial point alignment. Viability of the technique has been tested by processing recorded ECG data taken from six human patients.