Our results suggest that (1) none of the tested devices restore a comparable mechanical strength in the fractured specimens compared with the intact femurs, and (2) the "IT" possesses some biomechanical benefits for internal fixation of unstable femoral neck fractures compared with DHS and CS. Because the IT constructs failed with an inferior femoral neck fracture, complicating the mandatory anchorage of a prosthetic stem in a revision operation, more biomechanical experiments using the IT in the presence of a posterior comminution defect are required, along with clinical outcome studies.