The reduction of uncertainties for absolute piston gauge pressure measurements in the atmospheric pressure range

B.E. Welch, R. E. Edsinger, Vern E. Bean, Charles D. Ehrlich

Journal of Research of the National Institute of Standards and Technology · 1989 · 13 citations · 1 references

DOIFull text

Open access

Concepts

Abstract

NIST pressure calibration services with nitrogen are now based on two transfer standard piston gages for which the effective areas have been determined by calibration with the manometer developed at NIST for gas thermometry. Root-sum-squared three sigma uncertainties for the areas for the two gages are 3.05 ppm and 4.18 ppm.

References