An in vivo Evaluation of Different Methods of Working Length Determination

Nitin Shah, Sarita Singh, Jyoti Mandlik, Kalpana Pawar, Paras Gupta, Shoaib Ali Shaik

The Journal of Contemporary Dental Practice · 2013 · 24 citations · 21 references

DOIFull text

Open access

Concepts

Abstract

OBJECTIVE: The purpose of this in vivo study was to compare the ability of digital tactile, digital radiographic and electronic methods to determine reliability in locating the apical constriction. MATERIALS AND METHODS: Informed consent was obtained from patients scheduled for orthodontic extraction. The teeth were anesthetized, isolated and accessed. The canals were negotiated, pulp chamber and canals were irrigated and pulp was extirpated. The working length was then evaluated for each canal by digital tactile sensation, an electronic apex locator (The Root ZX) and digital radiography. The readings were then compared with post-extraction working length measurements. RESULTS: The percentage accuracy indicated that EAL method (Root ZX) shows maximum accuracy, i.e. 99.85% and digital tactile and digital radiographic method (DDR) showed 98.20 and 97.90% accuracy respectively. CLINICAL SIGNIFICANCE: Hence, it can be concluded that the EAL method (Root ZX) produced most reliable results for determining the accurate working length.