Journal of Orthopaedic Trauma · 2014 · 14 citations · 21 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryImplant SurgeryMechanical Behavior Of MaterialsStructural MaterialsBone MechanicsStrength PropertiesBendingBending Moment AnalysisBending StrengthFailure ModesInsertion Angle MeasurementMechanical Load EffectsScrew FixationScrew Fixation Mechanics
OBJECTIVES: The purposes of this study were to evaluate the cantilevered bending strength and failure modes of locking screws inserted at various angles in a plate with fully circumferential threaded holes. As an additional measure, the amount of screw head prominence at these angles was also assessed. METHODS: Standard 3.5-mm locking screws were inserted into round fully circumferential threaded holes through a standard straight 3.5-mm locking plate at various angles. The achieved angle of insertion and its prominence protruding from the far-bone side of the plate was measured using an optical luminescence technique. Each screw was then loaded at a constant rate until failure in a cantilevered bending scenario. The maximum cantilevered bending strength was measured, and the moment at failure was calculated. RESULTS: There was a positive correlation between increasing insertion angle and increasing prominence; a higher screw insertion angle yielded greater prominence. Prominence values ranged from negligible to 2 mm. As screw insertion angle increased, the bending moment at failure decreased. Screws inserted to 3 degrees or below primarily failed through screw deformation at the minor diameter below the head, whereas screws inserted to greater than 3 degrees primarily failed through locking mechanism disengagement. CONCLUSIONS: These findings indicate that cross threading may not be biomechanically advantageous and may change screw mode of failure. Based on these findings, screws inserted to 3 degrees or higher would reduce the bending moment at failure to approximately 50% of an orthogonally inserted screw.
21
Wade R. Smith, Bruce H. Ziran, Jeffrey O. Anglen et al. · Journal of Bone and Joint Surgery · 2007 · 166 citations
Tibial plateau fractures: Internal fixation with locking plates and the MIPO technique
Francesco Biggi, S. Di Fabio, C. D’Antimo et al. · Injury · 2010 · 125 citations