Interobserver Variability in the Measurement of Lower Leg Compartment Pressures

66

Citations

36

References

2015

Year

Abstract

Variations in use of a commercially available pressure monitor exist, and errors are common. Proper use improved accuracy, but even with proper technique, 40% of the measurements were >5 mm Hg from the actual pressure. Based on our data, measurement accuracy with this device should be questioned and viewed within a range. Regular review and education of technique is strongly recommended.