Publication | Open Access
The Relationship Between Oxygen Reserve Index and Arterial Partial Pressure of Oxygen During Surgery
105
Citations
8
References
2016
Year
These findings suggest that ORI >0.24 can distinguish PaO2 ≥100 mm Hg when SpO2 is over 98%. Similarly, ORI > 0.55 appears to be a threshold to distinguish PaO2 ≥150 mm Hg. The usefulness of these values should be evaluated prospectively. Decreases in ORI to near 0.24 may provide advance indication of falling PaO2 approaching 100 mm Hg when SpO2 is >98%. The clinical utility of interventions based on continuous ORI monitoring should be studied prospectively.
| Year | Citations | |
|---|---|---|