Concepedia

Publication | Closed Access

Comparative Biomechanical Analysis of Supracondylar Femur Fracture Fixation: Locked Intramedullary Nail Versus 95-Degree Angled Plate

62

Citations

26

References

1997

Year

Abstract

Based on the authors' results, the selection of a GSH nail or a DCS plate should not be determined by the severity of the fracture. If a DCS plate construct is selected, the authors recommend a dispersed screw configuration, including the most proximal hole in the plate, to provide superior stiffness in torsional loading and equal stiffness in axial loading when compared with the GSH nail constructs. If a GSH nail is selected, the authors recommend a grouped screw configuration, which absorbed more energy during axial loading compared with the DCS plate constructs and the nail with the dispersed screw configuration.

References

YearCitations

Page 1