Radiation Protection Dosimetry · 2014 · 12 citations · 7 references
In autumn 2010, an outdoor measuring station for measurement of atmospheric radon, gamma equivalent dose rate in the range of 100 nSv h(-1)-1 Sv h(-1) and proper meteorological parameters such as thermal air gradient, relative air humidity, wind speed and direction and solar radiation intensity was built in the area of the National Radiation Protection Institute vvi. The station was designed to be independent of an electrical network and enables on-line wireless transfer of all data. After introduction of the station, illustrations of its measurement properties and the results of measured diurnal and seasonal variability of atmospheric radon, based on annual continuous measurement using a high-volume scintillation cell at a height of 2.5 m above the ground, are presented.
7
222Radon as a Tracer of Atmospheric Motions: A Study in Milan
L. Sesana, L. Barbieri, U. Facchini et al. · Radiation Protection Dosimetry · 1998 · 27 citations
Agnieszka Podstawczyńska, Krzysztof Kozak, Włodzimierz Pawlak et al. · Nukleonika · 2010 · 27 citations
Environmental Radiochemistry, Earth Science, Environmental Monitoring +15