Measuring Cranial Vault Volume With Three-Dimensional Photography
73
Citations
0
References
2010
Year
Computed TomographyEngineeringCranial Vault VolumeSurgeryAnatomyDigital 3DGold StandardCt ScanRadiologyGeometric ModelingMedical ImagingNeuroimagingCraniofacial GrowthMedical Image ComputingVolume RenderingBone ImagingBiomedical Imaging3D ScanningCraniofacial SurgeryMedicineCraniofacial Disorder3D Imaging
The gold standard for measuring cranial vault volume is a computed tomography (CT) scan. Computed tomography scans in the pediatric population carry the significant risk of inducing a malignancy later in life. In a novel study, we show that a three-dimensional (3D) photograph can be used to measure and track cranial vault volume changes in craniosynostosis patients. In a study of adequate power as determined by pre hoc analysis, we demonstrate a greater than 0.91 correlation coefficient between volume as measured by CT and 3D photograph in more than 70 patients presenting to a tertiary craniofacial center (P < 0.001). Volume above a plane running through the lateral canthus (ex) and tragus (t) sag-ex-t on the digital 3D photograph differs from CT by a mean percentage difference of 31.9% (SD, 14.0%; 95% confidence interval, P < 0.0001). A linear regression model was used to determine the equation describing the relationship between volume as measured by 3D photograph and CT; with this equation, the absolute volume of the cranial vault is easily determined using only the 3D image. Interrater reliability is high (>0.99, ICC). The ability to measure cranial vault volume from the 3D photograph adds an important dimension to the complete and objective analysis of skull growth and postoperative change.