Position of TORP on the Stapes Footplate Assessed With Cone Beam Computed Tomography

Masahiro Komori, Naoaki Yanagihara, Jun Hyodo, Shinya Miuchi

Otology & Neurotology · 2012 · 15 citations · 15 references

Concepts

Abstract

OBJECTIVE: To improve hearing outcomes of ossiculoplasty with a total ossicular replacement prosthesis (TORP) in ears lacking a footplate superstructure, 3-dimensional computed tomography (3D CT) images were reconstructed to verify the position of the TORP on the footplate and contact between the TORP and the stapes footplate. STUDY DESIGN: Retrospective study. SETTING: University-affiliated tertiary referral center. PATIENTS: Six ears of 6 patients who underwent ossiculoplasty with TORP and were followed for greater than 3 years postoperatively. INTERVENTION: Cone beam CT (CBCT) images. MAIN OUTCOME MEASURES: An en face view of the stapes footplate and a volume-rendered 3D image were reconstructed. RESULTS: The en face view of the stapes footplate showed whether the TORP was centered on the stapes footplate. Volume-rendered 3D CBCT images revealed TORP malpositioning or migration, which were not detected on 2-dimensional CBCT images. In such cases, the TORP shaft was in contact with the wall of the oval window niche or the TORP had moved from the stapes footplate. CONCLUSION: Accurate visualization of TORP location on the footplate is important. Images that accurately show the position of the TORP on the stapes footplate will help improve hearing outcomes.

References