ASAIO Journal · 2014 · 18 citations · 6 references
Life SciencesMicrobiologyPediatricsBacteriologyPediatric Infectious DiseasesSepsisBacteremia ManagementBacterial InfectionsBloodstream InfectionsClinical BacteriologyExtracorporeal Membrane OxygenationPediatric SepsisAnaerobic Bacterial InfectionsAnaerobic Bloodstream InfectionsPathogen ReservoirsPediatric BacteremiaPediatric Extracorporeal Membrane Oxygenation
We report a case of bacteremia secondary to Cupriavidus pauculus in a 15-month-old boy on extracorporeal membrane oxygenation (ECMO). The source of the organism was water in the thermoregulator reservoir. The child responded well to cefepime and ciprofloxacin, a delayed oxygenator change out and replacement of the thermoregulator reservoir with a unit that was cleaned and decontaminated with sodium hypochlorite. Isolation of Cupriavidus pauculus from a patient on ECMO support should raise suspicion of the reservoir as a source.
6
Joan-Miquel Balada-Llasat, Camille T. Elkins, Lettie Swyers et al. · Journal of Clinical Microbiology · 2010 · 31 citations · Full text