Publication | Closed Access
Automated Fibrinogen Determination
36
Citations
0
References
1971
Year
Plasma FibrinogenEngineeringFibrinogen LevelDiagnosisPathologyBiomedical EngineeringThrombosisBioanalysisHematologyBiostatisticsAnalytical ChemistryClinical ChemistryBleeding DisorderAnalytical BiotechnologyLaboratory MedicineBiophysicsChromatographyFibrinolysisBiomedical AnalysisFibrin Split ProductsHemostasisCoagulopathyMedicine
A simple, rapid, and accurate semiautomated micromethod for the determination of plasma fibrinogen is described. One tenth milliliter of plasma is diluted in 1.9 ml. barbital buffer (pH 7.5). The clotting time is determined by an electronic tinier after addition of 0.1 ml. thrombin (100 U. per ml.) to 0.1 ml. diluted plasma. The fibrinogen level is read from a calibration curve. Abnormal levels are confirmed by a chemical method. Discrepancies between results obtained by the two methods may be a clue to structural abnormalities of fibrinogen or the presence of fibrin split products.