Journal of Thoracic Imaging · 2001 · 37 citations · 13 references
MedicineOncologyRadiologyCancer ScreeningDiagnostic RadiologyRadiation DetectionChest RadiographyComputed TomographyComputed Tomography Dose ManagementComputed Tomography PhysicsRadiation Dose MeasurementThoracic RadiographyComputed Tomography Dose MeasurementDose Reduction TechniquesFilter DesignHelical Computed TomographyLung Cancer ScreeningRadiation Dose ManagementRadiation Dose OptimizationRadiation Dose Reduction
A new aluminum filter, 5.8 mm thick at the center, was designed. The effective energy, exposure dose, absorbed dose, and noise were measured by using low-dose technique, very low-dose technique with a conventional filter, and very low-dose technique with a new filter on a chest phantom. Accuracy of very low-dose computed tomography (CT) with a new filter was compared against standard helical CT in 40 patients and against chest radiography in 35 patients. Effective energies were 42.6 keV and 51.6 keV at a conventional filter and the new filter, respectively. Compared against 20mA with a conventional filter, exposure dose was reduced by 17%, and absorbed dose was equivalent, at 30 mA with the new filter. Noise was improved by 9%. Compared with standard helical CT, the sensitivity, specificity, and accuracy of very-low-dose helical CT were 100%, 88%, and 95%, respectively. Very-low-dose helical CT was found to be significantly superior to chest radiography in the detection of lung cancers. Using a smaller tube current and an appropriate filter allows a further reduction in radiation dose in helical CT for lung cancer screening.
13
Peripheral lung cancer: screening and detection with low-dose spiral CT versus radiography.
Masahiro Kaneko, K. Eguchi, Hironobu Ohmatsu et al. · Radiology · 1996 · 841 citations
Low-dose CT of the lungs: preliminary observations.
David P. Naidich, Christopher H. Marshall, C Gribbin et al. · Radiology · 1990 · 338 citations
Medicine, Pediatrics, Radiology +16