Concepedia

Publication | Closed Access

The Accuracy, Precision and Reliability of Measuring Ventilatory Rate and Detecting Ventilatory Pause by Rainbow Acoustic Monitoring and Capnometry

87

Citations

15

References

2013

Year

Abstract

In this study of a population of postsurgical patients, the acoustic monitor and capnometer both reliably monitored ventilatory rate. The acoustic monitor was statistically more accurate and more precise than the capnometer, but differences in performance were modest. It is not known whether the observed differences are clinically significant. The acoustic monitor was more sensitive to detecting pauses in ventilation. Acoustic monitoring may provide an effective and convenient means of monitoring ventilatory rate in postsurgical patients.

References

YearCitations

Page 1