Accuracy of a dose map method assessed in clinical and anthropomorphic phantom situations using Gafchromic films

C. Bordier, R. Klausz, L. Desponds

Radiation Protection Dosimetry · 2015 · 21 citations · 9 references

Concepts

Abstract

A dose map method has been integrated on GE x-ray angiographic systems to provide an indication of the local dose distributed on a patient envelope representative of individual patient shapes. Tests have been performed to assess the accuracy of the method by using Gafchromic XR-RV3 films in an anthropomorphic phantom situation and in a clinical situation. Dose values inside different exposed areas have been compared between the film and the dose map method. The dose map results show a good visual agreement for the anthropomorphic phantom situation, and the local doses agreed within better than 15 % compared with the Gafchromic films in both situations.