Radiation Protection Dosimetry · 2015 · 21 citations · 9 references
MedicineBiomedical EngineeringRadiologyDiagnostic ImagingDiagnostic RadiologyRadiation DosimetryMedical ImagingPatient-specific DosimetryPhantom DosimetryRadiographic ImagingThree-dimensional Dose Distribution3D Dose MappingAnthropomorphic Phantom DosimetryGafchromic Film DosimetryImage-based Dose CalculationImaging-based DosimetryPhantom-based Dose MeasurementSpatial Dose Distribution AnalysisX-ray Imaging Systems
A dose map method has been integrated on GE x-ray angiographic systems to provide an indication of the local dose distributed on a patient envelope representative of individual patient shapes. Tests have been performed to assess the accuracy of the method by using Gafchromic XR-RV3 films in an anthropomorphic phantom situation and in a clinical situation. Dose values inside different exposed areas have been compared between the film and the dose map method. The dose map results show a good visual agreement for the anthropomorphic phantom situation, and the local doses agreed within better than 15 % compared with the Gafchromic films in both situations.
9
Radiation-induced skin injuries from fluoroscopy.
Thomas B. Shope · Radiographics · 1996 · 324 citations