Journal of Pediatric Orthopaedics · 1997 · 16 citations · 21 references
MedicineOrthopedic SurgerySurgeryCraniofacial SurgeryOrthotic TherapyOrthoticsExternal Skeletal FixationFixation BiomechanicsInternal FixationTrauma BiomechanicsBiomechanics Of Internal FixationCompressive LoadsDeflection MeasurementFixation Construct BiomechanicsFixation StiffnessMechanical Property EvaluationPaired T-testPins
Fixation of the halo device in the skulls of young children is sometimes compromised because of a thinner cortex and the increased jarring and trauma that small children experience during their everyday activities of life at that age. To assess the increased fixation that an eight-pin halo fixation would provide over the more standard four-pin fixation, the stiffness of fixation of a halo orthosis was evaluated in four dried human cadaver skulls. Two different fixation systems were used: the first one with four pins in the standard locations and the second with four additional pins. A compressive force was applied to the halo ring at a constant rate of 1.27 mm/min. Load and deflection data were recorded simultaneously until a force of 400 N was reached. Results for the two-pin systems were analyzed and compared statistically by using paired t tests. Statistically significant differences were established between the two sets of results with the 8-pin system providing an increase of 24.2% in the stiffness of fixation.
21
Complications in the use of the halo fixation device.
Steven R. Garfin, Michael J. Botte, R L Waters et al. · Journal of Bone and Joint Surgery · 1986 · 302 citations
Halo Application in the Infant
Scott J. Mubarak, Jon F. Camp, William Vuletich et al. · Journal of Pediatric Orthopaedics · 1989 · 105 citations