Analytical Chemistry · 2010 · 44 citations · 5 references
Natural SciencesChemistryMaterials ScienceAnalytical ChemistryMass Spectrometric AnalysisSpectroscopic AnalysisAnalytical Method ValidationGas ChromatographyInstrumental AnalysisMass SpectrometryPolymer AnalysisContraband DetectionDrop-on-demand PrintingExplosives AnalysisGas Chromatography-mass SpectrometryInkjet DepositionIon MobilityIon Mobility SpectrometryRdx
The feasibility of the use of piezoelectric drop-on-demand inkjet printing to prepare test materials for trace explosive analysis is demonstrated. RDX (1,3,5-trinitro-1,3,5 triazcyclohexane) was formulated into inkjet printable solutions and jetted onto substrates suitable for calibration of the ion mobility spectrometry (IMS) instruments currently deployed worldwide for contraband screening. Gravimetric analysis, gas chromatography/mass spectrometry (GC/MS), and ultraviolet-visible (UV-vis) absorption spectroscopy were used to verify inkjet printer solution concentrations and the quantity of explosive dispensed onto test materials. Reproducibility of the inkjet printing process for mass deposition of the explosive RDX (1,3,5-trinitro-1,3,5 triazcyclohexane) was determined to be better than 2% for a single day of printing and better than 3% day-to-day.
5
Instrumentation for trace detection of high explosives
David S. Moore · Review of Scientific Instruments · 2004 · 804 citations
Geoffrey B. Smith, Gary A. Eiceman, Margaret Walsh et al. · Field Analytical Chemistry & Technology · 1997 · 15 citations · Full text