Publication | Closed Access
A biomechanical comparison of static versus dynamic lag screw modes for cephalomedullary nails used to fix unstable peritrochanteric fractures
29
Citations
24
References
2011
Year
There is a 60 N/mm (12.4%) reduction in axial stiffness when the lag screw is in dynamic mode. Given the statistically significant reduction in axial and lateral stiffness with use of the dynamic mode, static lag screw mode should be further explored clinically for treatment of unstable peritrochanteric fractures.