Electrochemical Detection of Peroxynitrite Using a Biosensor Based on a Conducting Polymer−Manganese Ion Complex

Wei Choon Alvin Koh, Jung Ik Son, Eun Sang Choe, Yoon‐Bo Shim

Analytical Chemistry · 2010 · 49 citations · 27 references

Concepts

Abstract

A peroxynitrite (ONOO(-)) biosensor has been developed through the preparation of a new manganese-[poly-2,5-di-(2-thienyl)-1H-pyrrole)-1-(p-benzoic acid)] (Mn-pDPB) complex. DPB monomer was first synthesized and polymerized for the purpose of providing a polymer backbone for complex formation with Mn(2+) ion. The Mn-pDPB complex was characterized via Magnetomotive Force (MMF) simulation, X-ray photoelectron spectroscopy (XPS), and cyclic voltammetry. The complex selectively enhanced the reduction process of ONOO(-) which was used as the analytical signal for chronoamperometric detection. A polyethyleneimmine (PEI) layer was coated on the complex surface to increase selectivity and stability. The chronoamperometric calibration plot showed the hydrodynamic range of 2.0 × 10(-8)-5.0 × 10(-7) M. The detection limit was determined to be 1.9 (±0.2) × 10(-9) M based on S/N = 3. The microbiosensor, fabricated on a 100 μm diameter Pt tip, was applied in a real rat plasma sample for the detection of spiked concentrations of ONOO(-). The reliability and long-term stability of the microbiosensor was also examined with YPEN-1 cells in vitro, and the results shown were promising.

References

27