Journal of Research of the National Institute of Standards and Technology · 2004 · 17 citations · 4 references
EngineeringChemical EngineeringFluid DynamicsInstrumentationMetrologyFlow MeasurementFlow MetrologyMeasurement StandardsMeasurement UncertaintyPressure MeasurementPressure MetrologyCylinder Pressure MeasurementGas FlowGas Flow MetrologyIn-cylinder Pressure MeasurementPrimary Flow StandardsPrimary Pressure StandardsPrimary StandardsVolumetric Instrumentation
We describe two primary standards for gas flow in the range from 0.1 to 1000 μmol/s. (1 μmol/s ≅ 1.3 cm(3)/min at 0 °C and 1 atmosphere.) The first standard is a volumetric technique in which measurements of pressure, volume, temperature, and time are recorded while gas flows in or out of a stainless steel bellows at constant pressure. The second standard is a gravimetric technique. A small aluminum pressure cylinder supplies gas to a laminar flow meter, and the integrated throughput of the laminar flow meter is compared to the weight decrease of the cylinder. The two standards, which have standard uncertainties of 0.019 %, agree to within combined uncertainties with each other and with a third primary standard at NIST based on pressure measurements at constant volume.
4
Quartz capillary flow meter for gases
Robert F. Berg · Review of Scientific Instruments · 2004 · 29 citations