Radiation Protection Dosimetry · 2015 · 11 citations · 7 references
EngineeringNuclear EngineeringRadiation DetectionRadiation DosimetryRadiation MeasurementDosimetry InstrumentationRadiation DetectorsThermoluminescent DosimetryDetectorsDosimeter FadingFadingRoom-temperature StorageStorageThermal StabilityThermoluminescent DetectorsThermoluminescent Dosimeters
Thermoluminescence dosemeters are widely used in individual and environmental monitoring. The aim of this work was to compare the thermal stability of dosemeters of the Ext-Rad and whole-body card types with LiF:Mg,Ti and LiF:Mg,Cu,P detectors stored at different temperatures and periods. The dosemeters were stored at 0°C, room temperature and 40°C for periods that lasted 8, 30, 45, 90 and 120 d. In general, TLD-100H detectors present higher TL signal stability than TLD-100 detectors. The intensity of the signal remained constant for both materials for storage periods at 0°C. At RT the same results was observed for TLD-100H. For TLD-100 detectors, a maximum variation of 22 % was registered for the longest period. At 40°C the TL signal decreased with storage time for both detectors. The TL signal of TLD-100H detectors presented maximum variations of 12 % whereas for TLD-100 detectors, larger variations of 25 % were observed.
7
Lithium Fluoride: From LiF:Mg,Ti to LiF:Mg,Cu,P
P. Bilski · Radiation Protection Dosimetry · 2002 · 144 citations
Applications of Thermoluminescence Dosimetry in Medicine
Tomas Kron · Radiation Protection Dosimetry · 1999 · 119 citations