Major Upper Abdominal Surgery Alters the Calibration of Bioreactance Cardiac Output Readings, the NICOM, When Comparisons Are Made Against Suprasternal and Esophageal Doppler Intraoperatively

Li Huang, Lester A. H. Critchley, Jie Zhang

Anesthesia & Analgesia · 2015 · 30 citations · 17 references

Abstract

Doppler monitoring used in tandem provided valid trend lines of cardiac output changes against which NICOM readings could be compared. Intraoperatively, the NICOM was shown to track changes in cardiac output reliably in most circumstances. However, surgical interventions to the upper abdomen caused shifts in readings by >1 L/min/m, and the direction of the shifts was unpredictable. Anesthesiologists need to be aware of these calibration shifts and anticipate their occurrence, whenever the NICOM is used intraoperatively.