MedicineOncologyRadiation OncologyClinical Radiation OncologyDiagnostic ImagingTreatment PlanningImage-guided RadiotherapyImage-guided TherapyProton DosimetryProton TherapyRadiation Imaging SystemsRadiation TherapyBeam Delivery SystemsCommissioningComputed Tomography PhantomsImage-guided Proton TherapyProton Therapy DosimetryProton Therapy Phantom DosimetryTreatment Planning System ValidationTreatment System Validation
The purpose of this study is to describe the comprehensive commissioning process and initial clinical experience of the Mevion S250 proton therapy system, a gantry-mounted, single-room proton therapy platform clinically implemented in the S. Lee Kling Proton Therapy Center at Barnes-Jewish Hospital in St. Louis, MO, USA. The Mevion S250 system integrates a compact synchrocyclotron with a C-inner gantry, an image guidance system and a 6D robotic couch into a beam delivery platform. We present our commissioning process and initial clinical experience, including i) CT calibration; ii) beam data acquisition and machine characteristics; iii) dosimetric commissioning of the treatment planning system; iv) validation through the Imaging and Radiation Oncology Core credentialing process, including irradiations on the spine, prostate, brain, and lung phantoms; v) evaluation of localization accuracy of the image guidance system; and vi) initial clinical experience. Clinically, the system operates well and has provided an excellent platform for the treatment of diseases with protons.
12
A pencil beam algorithm for proton dose calculations
Linda Hong, Michael Goitein, M. Bucciolini et al. · Physics in Medicine and Biology · 1996 · 407 citations