IEEE Transactions on Medical Imaging · 2008 · 105 citations · 43 references
MedicineNuclear MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyComputed TomographyComputed Tomography PhysicsImagingRadiation Imaging SystemsRadiotracer LocalizationAdaptive AcquisitionAdaptive ObserversAdaptive SystemsPhoton EmissionRadiotracer Distribution PatternsRadiotracer QuantificationSingle-photon Emission
Adaptive imaging systems alter their data-acquisition configuration or protocol in response to the image information received. An adaptive pinhole single-photon emission computed tomography (SPECT) system might acquire an initial scout image to obtain preliminary information about the radiotracer distribution and then adjust the configuration or sizes of the pinholes, the magnifications, or the projection angles in order to improve performance. This paper briefly describes two small-animal SPECT systems that allow this flexibility and then presents a framework for evaluating adaptive systems in general, and adaptive SPECT systems in particular. The evaluation is in terms of the performance of linear observers on detection or estimation tasks. Expressions are derived for the ideal linear (Hotelling) observer and the ideal linear (Wiener) estimator with adaptive imaging. Detailed expressions for the performance figures of merit are given, and possible adaptation rules are discussed.
43
The Generalization of Student's Ratio
Harold Hotelling · The Annals of Mathematical Statistics · 1931 · 1.5K citations · Full text
Phase Retrieval And Diversity In Adaptive Optics
Robert A. Gonsalves · Optical Engineering · 1982 · 610 citations