Optics Express · 2016 · 51 citations · 17 references
Natural SciencesBiomedical EngineeringMaterials ScienceOptical MaterialsRadiation DosimetryRadiation MeasurementAbsorbed DoseFiber OpticsIn Vivo DosimetryOptical FibersRadiation Dose MeasurementScintillator MaterialsAngular DependenceDose RepeatabilityExternal Beam Radiotherapy DosimetryOptical Fiber WaveguidesPolymethyl MethacrylateRadiation IntensityScintillationScintillation Dosimetry
S:Tb), capable of emitting visible light at around 545 nm which is ideal for transmission through the PMMA when exposed to ionizing radiation was embedded in the PMMA plastic fiber. The dose rate of incident ionizing radiation is measured by analyzing the signal intensity emitted from the scintillation material which propagates through the fiber to a distal MPPC (multi-pixel photon counter). The dosimeter exhibits good repeatability with an excellent linear relationship between the fiber-optic dosimeter output and the absorbed radiation dose with an outstanding isotropic response in its radial angular dependence.
17