PubMed · 1977 · 79 citations · 4 references
MedicineNuclear MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyBrain ImagingMedical Imaging InstrumentationScintillation DetectorsCerebral Blood Flow QuantificationRadionuclide QuantificationRegional Cerebral Blood FlowScintillation DetectionSodium Iodide Detectors
A computer-based scintillation camera has been designed for both dynamic and static radionuclide studies. The detecting head has 254 independent sodium iodide crystals, each with a photomultiplier and amplifier. In dynamic measurements simultaneous events can be recorded, and 1 million total counts per second can be accommodated with less than 0.5% loss in any one channel. This corresponds to a calculated deadtime of 5 nsec. The multidetector camera is being used for 133Xe dynamic studies of regional cerebral blood flow in man and for 99mTc and 197 Hg static imaging of the brain.
4
Design and performance characteristics of a whole-body positron transaxial tomograph.
Hoffmann Ej, Phelps Me, Mullani Na et al. · PubMed · 1976 · 154 citations
Regional Cerebral Blood Flow in Focal Cortical Epilepsy
K Hougaard, Takanori Oikawa, E. Sveinsdottir et al. · Archives of Neurology · 1976 · 151 citations
Medicine, Neurology, Neuroscience +17