EngineeringBiomedical EngineeringComputer EngineeringSystem-on-chip DesignUltrasonographyUltrasoundDiagnostic Imaging SystemsDiagnostic UltrasonographyMedical ImagingMedical Imaging InstrumentationPoint-of-care UltrasoundUltrasound ImagingDelay CalculationPortable Imaging DevicesPortable Ultrasound SystemsPower ConsumptionSystem-on-chip ArchitectureUltrasonic ImagingUltrasound Imaging System DesignUltrasound Signal Processing Hardware
In this paper, we present a novel system-on-chip (SOC) solution for a portable ultrasound imaging system (PUS) for point-of-care applications. The PUS-SOC includes all of the signal processing modules (i.e., the transmit and dynamic receive beamformer modules, mid- and back-end processors, and color Doppler processors) as well as an efficient architecture for hardware-based imaging methods (e.g., dynamic delay calculation, multi-beamforming, and coded excitation and compression). The PUS-SOC was fabricated using a UMC 130-nm NAND process and has 16.8 GFLOPS of computing power with a total equivalent gate count of 12.1 million, which is comparable to a Pentium-4 CPU. The size and power consumption of the PUS-SOC are 27×27 mm(2) and 1.2 W, respectively. Based on the PUS-SOC, a prototype hand-held US imaging system was implemented. Phantom experiments demonstrated that the PUS-SOC can provide appropriate image quality for point-of-care applications with a compact PDA size ( 200×120×45 mm(3)) and 3 hours of battery life.
31
Elastography: A Quantitative Method for Imaging the Elasticity of Biological Tissues
J. Ophir, I. Céspedes, H. Ponnekanti et al. · Ultrasonic Imaging · 1991 · 3.5K citations