Concepedia

Publication | Open Access

Safe use of subdermal needles for intraoperative monitoring with MRI

15

Citations

6

References

2016

Year

Abstract

The authors were satisfied with the safety of their site-specific procedures and have begun with off-label use (following institutional review board approval and obtaining patient informed consent) of tested monitoring leads in cases that combine IOM and MRI. The authors recommend that all facilities perform their own site-specific testing of monitoring leads before proceeding with their routine use.

References

YearCitations