Publication | Open Access
Combined leadless pacemaker and subcutaneous implantable defibrillator therapy: feasibility, safety, and performance
88
Citations
15
References
2016
Year
Combined LP and S-ICD therapy appears feasible in all animal experiments (n = 2) and in one human subject. No interference in sensing and pacing during intrinsic and paced rhythm was noted in both animal and human subjects. However, induced arrhythmia testing was not performed in the patient. Defibrillation therapy did not seem to affect LP function. More data on safety and performance are needed.