2003 · 13 citations · 0 references
Natural SciencesBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyDielectric ResponseDielectricsMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMedical ImagingRadiofrequency ExcitationB1 Field Inhomogeneity CompensationComposite PulsesConductivityDielectric EffectsEcho Planar ImagingHigh-field Magnetic Resonance ImagingMagnetic Resonance Imaging AcquisitionPulse Sequence Design
Image quality in high field magnetic resonance imaging can significantly suffer from dielectric resonances. In samples with high dielectric constants the vacuum wavelength is shortened significantly and approaches at stronger fields such as 3T typical object dimensions. Even though object size and sample conductivity also play important roles, wave propagation often becomes non-uniform resulting in B1 field inhomogeneities. In this work it is shown that under certain conditions composite RF-pulses can be used to compensate for respective B1 inhomogeneities. Phantom studies with echo planar imaging at 3T exhibited an improved intensity homogeneity when using the proposed composite RF-excitation scheme.