Journal of Bone and Joint Surgery - British Volume · 1988 · 58 citations · 10 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryOrthopaedic BiomaterialsOrthopedic BiomechanicsOrthopedicsBiomechanics Of AnchorageDevice DesignFixation BiomechanicsImplant BiomechanicsLigament MechanicsAnchorage Strength AssessmentBiomechanical Testing Of ImplantsOrthopedic ImplantsTensile DeformationTensile Strain
This study aimed to compare the holding strength of various commercially-available anchorage devices for artificial ligaments, so that surgeons might make a reasoned choice. Tensile tests to failure were performed on screws, bollards, toggles and staples which had been implanted into cadaveric bones. The holding strength of all devices correlated significantly with the local thickness of cortical bone, so it is recommended that anchorages should be placed away from the joint line, into diaphyseal bone if possible. A new trans-cortical grommet was developed which, when used around an AO screw, had significantly greater holding strength than the other devices.
10
Function of the knee joint in various activities.
J. B. Morrison · PubMed · 1969 · 319 citations
The effect of biopsy-hole shape and size on bone strength
CR Clark, C. Morgan, DA Sonstegard et al. · Journal of Bone and Joint Surgery · 1977 · 169 citations